Print, 5 print – Ohaus E1RW60 Explorer Balances User Manual

Page 79

Advertising
background image

25

Mode

Units

Setup

Print

Enter

4.5 Print

The Print menu provides a number of options which can be turned ON or OFF. It contains eight submenus: Auto
Print,
feature which includes selection of Off, Continuous, Interval and on Stability, Inter, specifies time interval for
automatic output of displayed data, Stable data-only feature, Numeric only or full display data for output, GLPCont,
GLPTare
, Reference which prints reference weight value and Lock which enable you to program balance param-
eters and to lock the settings.

Setup

Use keys to change,

press Enter

Change

Procedure

• Press the Setup button, CAL is displayed.

• Press

or

button until PRINT is displayed.

• Press Enter button to continue.

• Press

or

button until either AUTOPRT,

INTER, STABLE, NUMERIC, GLPCONT, GLPTARE,
REFEREN, LOCK or EXIT is displayed.

• Press Enter button to continue.

• Press

or

button and select either menu

setting or ON or OFF.

• Press Enter button to save setting.

• Press

or

button to select next item or EXIT.

• Press Enter button to continue.

Auto Print Feature
When enabled, the Auto Print feature causes the balance to automatically output display data in one of three
ways: continuously, at user specified time intervals, or upon stability. Default settings are shown bold.

OFF

when set on turns off the auto print feature

Cont

when set on, outputs printed data continuously

Inter

provides a user specified printing interval

On Stb

provides printed data only when a stable reading is achieved

Interval
Can be set to provide a specified printing interval between 1 and 3600 seconds.

Print Stable Data Only
When set On, this feature permits only stable display data to be output. OFF is the default setting.

Print Numeric Data Only
When Numeric Data Only function is turned ON, this allows the balance to output numeric data only for RS232
output. OFF is the default setting.