3.5 Percent Weighing

Percent Weighing is

enabled only when Percent is turned ON in the Mode submenu under Setup. Percent

weighing permits you to place a reference load on the balance, then view other loads as a percentage of the
reference. The load you place on the pan as a reference may be displayed as any percentage you select from 5 %
to 100 % (in 1 % increments). One hundred percent does not necessarily have to represent the reference load.
Subsequent loads, displayed as a percentage of the reference are limited only by the capacity of the balance.
The default setting is Reference 100 %. Refer to paragraph 4.8 to enable percent weighing.

Stable

%

Mode

Use keys to change,

press Enter

Change

Mode

Use keys to change,

press Enter

Change

%

Procedure

• Press the Mode button.

• Press

or

button until PERCENT is dis-

played.

• Press Enter button, PUT>PAN 100 % is displayed.

If a container is used, the balance can be tared at
this point. The % display momentarily blanks while
the balance is taring out.

• Put the reference load on the pan.

• Press

or

button and select reference

weight percentage (Percent Range 5 to 100). Hold

button down for fast change.

• Press Enter button to save setting, WORKING is

displayed ... calculating reference weight.

Balance displays reference weight for five seconds in
selected measuring unit, then displays the percentage.

• Remove the reference weight from the pan and

replace it with another load. The second load is dis-
played as a percentage of the reference.

NOTE: The PERCENT display (number of digits) is a
function of the accuracy of the balance and the size of
the reference weight. The display examples were with a
300 g mass used with an 8.1 kilogram balance.