4 trioxmatic® 700 iq (sw) setting table, Trioxmatic, Sensor (see section 3.4 – YSI IQ S ENSOR N ET TriOxmatic 700 IQ (SW) User Manual

Page 14

Advertising
background image

Commissioning

TriOxmatic

®

700 IQ (SW)

3 - 4

ba76013e01

01/2012

3.4

TriOxmatic

®

700 IQ (SW) setting table

Setting

Selection/values

Explanation

Measuring mode

Concentration
Saturation

Unit of the measured value on the measured
value display.

Measuring range
Concentration

0 ... 60.0 mg/l
0 ... 60.0 ppm

These measuring ranges are available for
selection.

Measuring range
Saturation

0 ... 600 %

The measuring range is set permanently.

Calibration

(software version 2.21
or higher)

valid
invalid

last valid

active

abort

Displays and determines which calibration
data the measured value calculation is
based on. The active calibration is displayed
in the calibration history (see section 4.2.3).

valid indicates that a valid calibration is
available. The value cannot be changed.

invalid is displayed if the last calibration is
invalid and the sensor is blocked for measu-
rement. In this case, you can switch to the
values of the last valid calibration, provided a
valid calibration is available in the sensor.
Thus you activate the last valid calibration
stored in the sensor the next time you exit the
setting table with Save and quit. The next
time the setting table is opened, valid is dis-
played.

active indicates that the sensor is being cali-
brated. If Cancel is selected, the active calib-
ration procedure is canceled as soon as the
setting table is exited with Save and quit.

Temperature mode

°C
°F

Unit of the measured temperature value
(Celsius, Fahrenheit).

Salinity

On
Off

Determines whether the entered salinity
should be taken into account.

Advertising