YSI IQ S ENSOR N ET TriOxmatic 700 IQ (SW) User Manual

Page 39

Advertising
background image

TriOxmatic

®

700 IQ (SW)

Maintenance, cleaning, disposal, and replacement

5 - 17

ba76013e01

01/2012

Cleaning the silver

counter electrode in a

beaker

Caution

The reference electrode must not come into contact with the
cleaning solution under any circumstances. This could destroy
the reference electrode and cause the sensor to become
defective.

16

Recalibrate the sensor (see section 4.2 C

ALIBRATION

).

1

Wipe the silver counter electrode with a lint free paper towel
and carefully remove any loose deposits.

2

Fill a beaker (150 ml, high-sided form) with approx. 25 ml RL-
AG/Oxi cleaning solution.

Advertising