Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
700 IQ (SW)
5 - 16
ba76013e01
01/2012
10
Rinse the sensor head and electrode unit several times with
deionized water.
11
Water the sensor head and electrode unit in deionized water
for at least an hour.
12
Carefully shake off the drops of water.
13
Fill a new WP 600 membrane cap and screw it on (see section
5.3 C
).
14
Connect the sensor to the sensor connection cable.
15
After approx. 60 minutes, the sensor is ready for operation.