ba76013e01

01/2012

9

Indexes

9.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S

ENSOR

N

ET

system for the TriOxmatic

®

700 IQ (SW) sensor.

Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Note
All message codes of the TriOxmatic

®

700 IQ (SW) end up with "331".

9.1.1

Error messages

Message code

Message text

EA1331

Meas. range exceeded or undercut
* Check process
* Select other meas. range

EA2331

Sensor temperature too high!
* Check process and application

EA3331

Sensor temperature too low!
* Check process and application

EC4331

Sensor could not be calibrated,

sensor blocked for measurement

Cause: instable signal
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)

EC5331

Sensor could not be calibrated,

sensor blocked for measurement

Cause: slope error
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)