2 cleaning the sensor shaft and membrane, Cleaning the sensor shaft and membrane -2, Section 5.2 c – YSI IQ S ENSOR N ET TriOxmatic 700 IQ (SW) User Manual

Page 24: Leaning, Sensor, Shaft, Membrane, Nsor (section 5.2

Advertising
background image

Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

700 IQ (SW)

5 - 2

ba76013e01

01/2012

5.2

Cleaning the sensor shaft and membrane

For normal operation (e.g. municipal wastewater), cleaning and
calibration are strongly recommended:

 in the case of pollution (according to visual check)

 in the case of suspected incorrect measured values (usually too low)

 if the measured value lies outside the range of precision required by

the user during the function check.

Cleaning agents

Cleaning

Caution

Acetic acid irritates the eyes and the skin. When handling acetic
acid, always wear protective gloves and protective goggles.

Contamination

Cleaning agents

For sludge and loosely
adhering dirt, or biological
deposits

Soft cloth or soft sponge, warm
tapwater with detergent

Salt and / or lime deposits

Acetic acid (volume percentage =
20 %),

soft cloth or soft sponge

1

Pull the sensor out of the sample.

2

Get rid of any coarse contamination on the sensor (e.g. brush
it off in a bucket of tapwater, wash it down with a hose or wash
it off with a cloth).

3

Unscrew the protective hood from the sensor.