Maintenance, cleaning, disposal, and replacement
5 - 19
ba76013e01
01/2012
7
Carefully shake off the drops of water.
8
Fill a new WP 600 membrane cap and screw it on (see section
5.3 C
).
9
Connect the sensor to the sensor connection cable.
10
After approx. 60 minutes polarization time, the sensor is ready
for operation.
11
Recalibrate the sensor (see section 4.2 C
).