Cleaning the silver counter electrode

We recommend to clean the silver counter electrode with the RA 600
cleaning attachment that is available as an accessory (see section
5.8 M

AINTENANCE

EQUIPMENT

AND

REPLACEMENT

PARTS

). This prevents

the reference electrode from coming into contact with the cleaning
solution. However, it is also possible to clean the silver counter
electrode in a beaker.

Caution

The reference electrode must not come into contact with the
cleaning solution under any circumstances. This could destroy
the reference electrode and cause the sensor to become
defective. However, no danger can arise if the RA 600 is used
correctly.

Cleaning the silver

counter electrode with

the RA 600

cleaning attachment

Caution

Never grease or lubricate the inner O-ring in the RA 600 cleaning
attachment!

1

Wipe the silver counter electrode with a lint free paper towel
and carefully remove any loose deposits.

2

Screw the cleaning attachment onto the sensor, instead of the
membrane cap.