Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

700 IQ (SW)

5 - 18

ba76013e01

01/2012

3

Clamp the sensor in a stand.

Submerse the electrode unit in the RL-AG/Oxi cleaning
solution to just above the silver counter electrode.

4

Leave the cleaning solution to work for 1 hour.

5

Rinse the sensor head and electrode unit several times with
deionized water.

6

Water the sensor head and electrode unit in deionized water
for at least an hour.