9.1.2

Info messages

EI1331

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

EI2331

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components,

replace components

ES1331

Component hardware defective
* Contact service

ESB331

SensReg: Electrolyte supply is depleted
* Change electrolyte solution and membrane head immediately (see
operating manual)

ESC331

SensLeck: Membrane head damaged
* Change electrolyte solution and membrane head immediately (see
operating manual)

Note: This piece of information is transmitted with the
TriOxmatic

®

700 IQ only.

Message code

Message text

Message code

Message text

IC1331

Sensor has been successfully calibrated
* For calibration data, see calibration history

IC4331

Last valid calibration has been activated. Make sure the sensor
operates correctly.

II1331

Language not available,

Default language German
* Contact service