Oxi 1970i

Operation

19

ba75691e01

01/2007

4.2.5

AutoRead AR (drift control) and hold function

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values. With the aid of the hold function the
measured value display is frozen.

Criteria

With identical measurement conditions, the following criteria are valid
for the AutoRead function:

Use the AutoRead function and hold function like this with the sensors

z Oxygen depth armature
z CellOx 325
z DurOx 325

Note
The special features of the self-stirring D. O. sensor StirrOx G can be
found in section 4.2.6 on page 20.

Measuring mode

Reproducibility

Response time

D. O. concentra-
tion

better than 0.05 mg/l

> 10 seconds

D. O. saturation
index

better than 0.6 %

> 10 seconds

1

Select the measuring mode with <M>.

2

Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.

4

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.