4.2.5
AutoRead AR (drift control) and hold function
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
Criteria
With identical measurement conditions, the following criteria are valid
for the AutoRead function:
Use the AutoRead function and hold function like this with the sensors
z Oxygen depth armature
z CellOx 325
z DurOx 325
Note
The special features of the self-stirring D. O. sensor StirrOx G can be
found in section 4.2.6 on page 20.
Measuring mode
Reproducibility
Response time
D. O. concentra-
tion
better than 0.05 mg/l
> 10 seconds
D. O. saturation
index
better than 0.6 %
> 10 seconds
1
Select the measuring mode with <M>.
2
Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
4
If necessary, start the next AutoRead measurement with
<RUN/ENTER>.