Oxi 1970i
What to do if...
49
ba75691e01
01/2007
6
What to do if...
Display LoBat
Instrument does not
react to keystroke
Error message
0FL
0FL
0FL
0FL
Error message
E3
E3
E3
E3
Cause
Remedy
– Battery almost empty
– Charge the battery (see
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <RCL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause
Remedy
Display range exceeded
Oxygen sensor:
– Not connected
– Connect the sensor
– Cable broken
– Replace sensor
– Depleted
– Replace sensor
– Short-circuit between gold and
lead electrode
– Clean sensor and replace it if
necessary
Cause
Remedy
Invalid calibration
Oxygen sensor:
– Electrolyte solution depleted
– Regenerate sensor
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate sensor
– Obsolete
– Replace sensor
– Broken
– Replace sensor