4.3

Calibration

Why calibrate?

D. O. sensors age. This changes the slope of the D. O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.
Thus, you should calibrate at regular intervals.

Calibration data records

The Oxi 1970i administrates two sets of calibration data:

Sensors of different classes can be calibrated separately. When a sen-
sor of one class is calibrated the calibration data of the other class re-
main stored. The Oxi 1970i recognizes the class of the connected
sensor and automatically uses the correct calibration data.

Calibration procedures

The Oxi 1970i provides 2 calibration procedures:

z Calibration in water vapor-saturated air.

To calibrate the CellOx 325, use theOxiCal

®

-SL air calibration ves-

sel (accessory), to calibrate the DurOx 325, use the OxiCal

®

-D air

calibration vessel (contained in the scope of delivery of the sensor).

z Calibration via a comparison measurement (e. g. Winkler titration

according to DIN EN 25813 or ISO 5813). At the same time, the rel-
ative slope is adapted to the comparison measurement by a correc-
tion factor.

When to calibrate?

z After connecting another D. O. sensor
z If the sensor symbol flashes, i. e. after the calibration interval has ex-

pired

Calibration interval

The calibration interval (Int 3) is set to 14 days in the factory. You can
select the interval in the range of 1 ... 999 days.

AutoRead

The calibration procedure automatically activates the AutoRead func-
tion.

Displaying calibration

data

Each time the instrument is switched on, the calibration data are shown
on the display for a short time (see section 4.1 S

WITCHING

ON

THE

MEA

-

SURING

INSTRUMENT

). In order to view the calibration data, switch the

measuring instrument off and switch it on again.

z Set 1, for sensor class 1:

– Oxygen depth armature

– CellOx 325, or

– StirrOx G

z Set 2, for sensor class 2:

– DurOx 325