3 pc/external printer (rs232 interface), 4 remote control, Pc/external printer (rs232 interface) – Xylem Oxi 1970i PORTABLE DISSOLVED OXYGEN METERS User Manual

Page 40: Remote control, Section 4.5.3, 1 cts 2 rxd 3 ground 4 txd

Advertising
background image

Operation

Oxi 1970i

40

ba75691e01

01/2007

4.5.3

PC/external printer (RS232 interface)

Via the RS 232 interface, you can transmit the data to a PC or an ex-
ternal printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the
interface to the devices.
The data output automatically switches to the RS232 interface.

Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.

Set up the following transmission data at the PC/printer:

Socket assignment

4.5.4

Remote control

The measuring instrument can be remotely controlled from a PC. This
requires the KOM pilot communication kit. It is available as an acces-
sory.
The instrument is then controlled via commands that simulate key-
strokes and request the current display contents.

Note
A more detailed description is provided within the scope of delivery of
the communication kit.

Baud rate

selectable between: 1200, 2400, 4800, 9600

Handshake

RTS/CTS + Xon/Xoff

Parity

none

Data bits

8

Stop bits

1

1 CTS
2 RxD
3 Ground
4 TxD

1

2

3

4

RS 232

REC