Er f – Xylem Oxi 1970i PORTABLE DISSOLVED OXYGEN METERS User Manual
Page 31

Oxi 1970i
Operation
31
ba75691e01
01/2007
Note
The AutoStore function is interrupted if you start other functions, e.g.
output the data storage.
After the function is finished, the AutoStore function is continued. By
this, however, temporal gaps in the recording of the measured values
will occur.
Switching off AutoStore
Switch AutoStore off by:
z setting the save interval (Int 1) to OFF, or
z switching the measuring instrument off and then on again.
6
Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
7
Set the required ID number with <
▲> <▼>.
8
Confirm with <RUN/ENTER>.
The instrument switches to the Oxi measuring mode and starts
the measuring and saving process.
AutoStore flashes on the display.