Xylem Oxi 1970i PORTABLE DISSOLVED OXYGEN METERS User Manual
Page 50

What to do if...
Oxi 1970i
50
ba75691e01
01/2007
Error message
E7
E7
E7
E7
AR flashes
continuously
Measured value too low
Measured value too
high
Display
to
to
to
to
Sensor symbol flashes
Cause
Remedy
Membrane damaged
– Membrane damaged
– Regenerate sensor
– Membrane head not screwed
on tight enough
– Screw membrane head tight
Cause
Remedy
No stable measured value
– Membrane contaminated
– Clean membrane
Cause
Remedy
– Insufficient
flow
– Provide flow to the sensor
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function
using the salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the
membrane
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
– Time-out of the interface
– Check the instrument
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system