3 entering the salt content (salinity), Entering the salt content (salinity) – Xylem Oxi 1970i PORTABLE DISSOLVED OXYGEN METERS User Manual
Page 27

Oxi 1970i
Operation
27
ba75691e01
01/2007
4.3.3
Entering the salt content (salinity)
A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent (the measured salinity) of the
test sample (range 0.0 - 70.0) and to switch on the salinity correction.
Entering the
salt content
Note
How to switch on the salt content correction is described on page 17.
Parameter
Value range
Salinity
0.0 ... 70.0 in steps of 0.1
1
Determining the salinity of the test sample (any method).
2
Press the <CAL> key repeatedly until Sal appears on the dis-
play.
3
Enter the salt content with <
▲> <▼>.
4
Switch to the measuring mode with <M>.
REL1
TP
Oxi
Cal
AR
°
C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
cm
1/
K
/
%
RCL
REL2
pH
/
mV
m
cm
S/
%
5
6 9 6
5 0
O
mg/ l
Sal
ARng