4.6
Configuration
You can adapt the measuring instrument to your individual require-
ments. To do this, the following parameters can be changed (the status
on delivery is marked in bold):
Note
You can leave the configuration menu at any time with <M> . The pa-
rameters that have already been changed are stored.
Baud rate
Baud rate
1200, 2400, 4800, 9600
Air pressure display
Current value in mbar
(no input possible)
Calibration interval (Int 3)
1 ... 14 ... 999 d
AutoRange ARng
On or off
Date/time
Any
1
Switch off the measuring instrument.
2
Press the <M> key and hold it down.
3
Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
4
Select the required baud rate with <
▲> <▼>.
5
Confirm with <RUN/ENTER>. On the display, the current air
pressure in mbar appears.