4.6

Configuration

You can adapt the measuring instrument to your individual require-
ments. To do this, the following parameters can be changed (the status
on delivery is marked in bold):

Note
You can leave the configuration menu at any time with <M> . The pa-
rameters that have already been changed are stored.

Baud rate

Baud rate

1200, 2400, 4800, 9600

Air pressure display

Current value in mbar
(no input possible)

Calibration interval (Int 3)

1 ... 14 ... 999 d

AutoRange ARng

On or off

Date/time

Any

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.

4

Select the required baud rate with <

> <>.

5

Confirm with <RUN/ENTER>. On the display, the current air
pressure in mbar appears.