4.3.2

Calibrating via a comparison measurement

Proceed as follows to calibrate the instrument:

1

Connect the D. O. sensor to the measuring instrument.

2

Press the <CAL> key repeatedly until O

2

FAC appears.

3

Immerse the D. O. sensor in the calibrating solution.

4

Press the <RUN/ENTER> key.
The AutoRead measurement begins. If the measured value is
stable, the instrument displays the determined measured value
and the correction factor of the relative slope (initial value =
1.000).

5

Remove the D. O. sensor from the calibrating solution.

6

Perform a comparison measurement, e. g. Winkler titration,
with the calibrating solution (determine the nominal value).
Leave the measuring instrument switched on until the nominal
value is determined.

7

Set the displayed value on the measuring instrument to the
nominal value with <

> <>. This changes the related correc-

tion factor of the relative slope.

REL1

S

AR

Year

Ident

No.

Day.Month

Tref25

Tref20

LoBat

Lin

nLF

mg/ l

ARng

Sal

pH

1

TEC

Auto

Store

8

Time

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

/

mV

m

cm

S/

%

F

1 7 6

A C

°

C

TP

O

Oxi

Cal

TP

REL1

S

Year

Ident

No.

Day.Month

Tref25

Tref20

LoBat

Lin

nLF

Sal

pH

1

TEC

Auto

Store

Time

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

/

mV

m

cm

S/

%

4

0 0 0

4 2

°

C

O

1

AR

ARng

Oxi

Cal

mg/ l

Advertising