Technical data

Oxi 1970i

56

ba75691e01

01/2007

Guidelines

and norms used

7.2

Measuring ranges, resolutions, accuracies

Measuring ranges,

resolutions

Note: The values given in brackets apply to the sensor DurOx 325 es-
pecially.

Accuracy

(± 1 digit)

Correction functions

EMC

E.C. guideline 89/336/EEC
EN 61326-1:1997
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A

Instrument safety

E.C. guideline 73/23/EEC
EN 61010-1 A2:1995

Climatic class

VDI/VDE 3540

IP protection

EN 60529:1991

Variable

Measuring range

Resolution

Concentration [mg/l]

0 ... 19.99 (0 ... 19.9)
0 ... 90.0 (0 ... 90)

0.01 (0.1)
0.1 (1)

Saturation [%]

0 ... 199.9 (0 ... 199)
0 ... 600

0.1 (1)
1

T [°C]

0 ... 50.0

0.1

Variable

Accuracy

Concentration [mg/l]

± 0.5 % of measured value
at ambient temperature of + 5 °C ... + 30 °C

Saturation [%]

± 0.5 % of measured value
when measuring in the range of ± 10 K
around the calibration temperature

T [°C]

± 0.1

Temperature com-
pensation

Accuracy better than 2 % at 0 ... + 40 °C

Salinity correction

0 ... 70.0 SAL