ProLab 1000

Operation

33

ba75576e04

01/2011

4.5

pH value / ORP voltage

4.5.1

General information

You can measure the following variables:

z pH value [ ]
z ORP [mV]

CAUTION

When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result. The
RS232 and USB interfaces are not galvanically isolated.

Temperature

measurement

For reproducible pH measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:

z Automatic measurement of the temperature by the temperature

sensor (NTC30 or Pt1000) integrated in electrode.

z Measurement by an external temperature sensor.
z Manual determination and input of the temperature.

The display of the temperature indicates the active temperature
measuring mode:

Preparatory activities

Perform the following preparatory activities when you want to measure:

Temperature

sensor

Resolution of the

temp. display

Mode

yes

0.1 °C

Automatic
with temperature sensor

-

1 °C

Manual

1

Connect a pH or ORP electrode to the meter.

2

If necessary, select the pH or mV display with <MODE>.

3

Adjust the temperature of the solutions and measure the
current temperature if the measurement is made without a
temperature sensor.

4

Calibrate or check the meter with the electrode.