6 what to do if, 1 ph and orp measurement, What to do if – Xylem ProLab 1000 BENCH TOP pH METER User Manual

Page 79: Ph and orp measurement, Chapter 6 w, 6what to do if

Advertising
background image

ProLab 1000

What to do if...

79

ba75576e04

01/2011

6

What to do if...

6.1

pH and ORP measurement

Error message

Err1

Error message

Err2

Error message

Err4

Error message

CalError

Cause

Remedy

pH electrode:

– Measured value outside the

measuring range

– Use a suitable electrode

– Air bubble in front of the

junction

– Remove air bubble

– Air in the junction

– Extract air or moisten

junction

– Cable broken

– Replace electrode

– Gel electrolyte dried out

– Replace electrode

Cause

Remedy

– No electrode connected

– Connect electrode

– Setting time during calibration

too long

– Adjust temperature if

necessary

– Recalibrate

Cause

Remedy

– Temperature not stable during

calibration.

– Adjust temperature if

necessary

– Recalibrate

Cause

Remedy

pH electrode:

– The values determined for zero

point and slope of the electrode
are outside the allowed limits.

– Recalibrate

– Junction contaminated

– Clean junction

– Electrode broken

– Replace electrode