2 automatic storage at intervals, Automatic storage at intervals – Xylem ProLab 1000 BENCH TOP pH METER User Manual

Page 63

Advertising
background image

ProLab 1000

Operation

63

ba75576e04

01/2011

4.7.2

Automatic storage at intervals

The storage interval (Interval) determines the chronological interval
between automatic storage processes. Each storage process transmits
the current dataset to the interface at the same time.

Configuring the

automatic storage

function

Settings

You can configure the automatic storage function with the following
settings:

1

Open the menu for automatic storage with <STO

_

>.

SAMPLE ID

1

Interval

30 s

Duration

180 min

Continue

00d03h00min

00d15h

34min

<STO_>

Menu item

Possible
setting

Description

SAMPLE ID

1 ... 999

Determine the sample
designation for the series
of datasets.

Interval

1, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min,
30 min, 60 min

Storage interval.

The lower limit of the
storage interval can be
restricted by the number of
free storage locations.
The upper limit is restricted
by the storage duration.

Duration

1 min ... x min

Storage duration.
Specifies after which time
the automatic storage
should be terminated.

The lower limit of the
storage duration is
restricted by the storage
interval.
The upper limit is restricted
by the number of free
storage locations.