ProLab 1000

What to do if...

81

ba75576e04

01/2011

CalClock flashes

6.2

ISE measurement

Error message Err1

Obviously incorrect

measured values

Error message, Error

(invalid

calibration)

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Measuring range exceeded

– Dilute test sample

Cause

Remedy

– Electrode not connected

– Connect electrode

– Cable broken

– Replace electrode

Cause

Remedy

ISE electrode:

– Moisture in the plug

– Dry plug

– Electrode obsolete

– Replace electrode

– Electrode unsuitable for the

range to be measured

– Use suitable electrode

– Socket damp

– Dry socket

Calibration procedure:

– Wrong sequence of standards

for three point calibration

– Select correct sequence

– Calibration standards do not

have the correct temperature
(max.

±

2 °C temperature

difference)

– Adjust the temperature of

the calibration standards