3 options for data transmission, Options for data transmission – Xylem ProLab 1000 BENCH TOP pH METER User Manual

Page 73

Advertising
background image

ProLab 1000

Operation

73

ba75576e04

01/2011

4.8.3

Options for data transmission

Via the RS232 interface, you can transmit data to a PC or external
printer. The following table shows which data is transmitted to the
interface in which way:

Note
The following rule applies: Except for the menus, the display content is
generally downloaded to the interface with <PRINT> (displayed
measured values, measurement datasets, calibration records).

Data

Control

Operation / description

Current
measured values
of the connected
sensor

Manual

z With <PRINT>.
z Simultaneously with every

manual storage process (see
page 62).

automatic, at
intervals

z With <PRT

_

>. Then you

can set the transmission
interval.

z Simultaneously with every

automatic storage process
(see page 63).

Stored measured
values

Manual

z Displayed dataset:

with <PRINT> after recall
from data storage.

z All datasets according to the

filter criteria via the Output to
RS232/USB
function.

For details, see page 65.

Calibration
records

Manual

z Calibration record of a sensor

with <PRINT> (after calling
up from the data storage or at
the end of a calibration
procedure).

z after calling up from the data

storage via the Output to
RS232/USB
function.

For details, see page 68.

automatic

z For the respective sensor at

the end of a calibration
procedure.