background image

Operation

ProLab 1000

38

ba75576e04

01/2011

4.5.4

Settings for pH and ORP measurements

Overview

The following settings are possible for pH and ORP measurements:

z Resolution
z Calibration interval
z Buffers for calibration
z Unit for zero point
z Unit for slope
z Calibration record (display)
z Calibration history

Settings

The settings are made in the measuring menu of the pH/ORP
measurement. To open it, activate the relevant measuring window in
the measured value display and press <MENU/OK>. After completing
the settings, switch to the measured value display with <MODE>.

Menu item

Possible setting

Description

Calibration /
Calibration record

-

Displays the calibration
record of the last
calibration.

Calibration /
Calibration history

-

Displays the calibration
history of the last
calibrations.

Calibration /Buffer

NIST/DIN
TEC
...

Buffer sets to be used for
pH calibration (see
page 40).

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing CalClock in the
measuring window.

Calibration /Unit for
zero point

mV
pH

Unit for the zero point.

Calibration /Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to
the Nernst slope of
-59.16 mV/pH
([determined slope/Nernst
slope] x 100).