Operation

31

8

Thoroughly rinse the electrode with deionized water.

9

If required, enter the temperature of the second buffer solution
with <

> <> while keeping the <RUN/ENTER> key de-

pressed.

10

Immerse the pH electrode in the second buffer solution.

11

Press the <RUN/ENTER> key.

12

Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <

> <>.

13

When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.

14

Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.

/