9 reset, Reset – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual
Page 58

Operation
pH/ 3400i
58
4.9
Reset
You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
Measurement
parameters
The following measuring parameters are reset to the delivery status:
pH
Oxi
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery
status:
Measuring mode
pH
Slope -59.16 mV/pH
Asymmetry
0 mV
Manual temperature input
25 °C
Measuring mode
D. O. concentration
Relative slope
1.00
Salinity (value)
0.0
Salinity (function)
off
Baud rate
4800
Display during the
pH calibration
Buffer nominal value
Interval 1
(automatic save)
OFF
Interval 2
(for data transmission)
OFF