4 dissolved oxygen, 1 general information, Dissolved oxygen 4.4.1 – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 33: General information

Advertising
background image

pH/ 3400i

Operation

33

4.4

Dissolved oxygen

4.4.1

General information

Note
D. O. measurements with the pH/ Oxi3400i can only be carried out us-
ing a CellOx 325 or StirrOx G D. O. probe. The stirrer of the StirrOx G
D. O. probe has to be supplied with voltage separately using the
NT/pH Mix 540 power supply.

You can measure the following variables:

z D. O. concentration
z D. O. saturation

The pH/Oxi 3400i handheld meter is equipped with the following func-
tions:

z AutoRange (automatic measurement range selection),
z The AutoRead function (drift control) for checking the stability of the

measurement signal. This ensures the reproducibility of the mea-
surement signal. For details of how to switch the AutoRead function
on/off, see page 36.

Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.

Preparatory activities

Perform the following preparatory activities when you want to measure:

Note
Incorrect calibration of D. O. probes will result in incorrect measured
values. Calibrate at regular intervals.

1

Connect the D. O. probe to the measuring instrument. The in-
strument automatically switches over to D. O. measurement
(O

2

status display).

If the D. O. probe is already connected, press the <M> repeat-
edly until the O

2

status display appears.

2

Calibrate or check the measuring instrument with the probe.
How to calibrate is described in section 4.4.5 from page 37.

3

Immerse the D. O. probe in the test sample.

4

Use <

> <> to toggle between the measuring modes, D. O.

concentration (mg/L) and D. O. saturation (%).