A l c – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 39

Advertising
background image

pH/ 3400i

Operation

39

3

Put the D. O. probe into the air calibration vessel.

4

Press the <CAL> key repeatedly until the calibration mode ap-
pears. The sensor symbol displays the evaluation of the last
calibration (or no sensor symbol in the delivery state or after the
measurement parameter has been reset).

5

Press the <RUN/ENTER> key. AutoRead is active, AR flashes.

6

As soon as a stable value is reached, the AR display indicator
stops flashing. The calibration is finished now. The value of the
relative slope appears on the display. The probe symbol shows
the probe evaluation (see page 38).

7

Switch to the measuring mode with <M>.

%

Auto

ARng

Sal

mg/l

cm

M
cm

8

LoBat

Time

Baud

Day.Month No.

Ident

Year

mbar

1

S

RCL

AR

Store

°

F

1/

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

6

7

1

A L

C

O

°

C

TP

Cal

mg/l

8

LoBat

Time

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

1

1/cm

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

6

7

1

8 8

0

O

Cal

AR

S

ARng

mg/l

8

LoBat

Time

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

1

RCL

Store

Auto

°

F

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

6

7

1

8 8

0

O

°

C

TP

Cal

AR

S

Advertising