C a l – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 50

Advertising
background image

Operation

pH/ 3400i

50

Outputting the calibra-

tion protocol on the

interface

Sample printout:

1

Press the <RCL> key repeatedly until CAL SEr appears on the
display.

2

Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmit-
ted to the interface. After the data transmission, the instrument
automatically switches to the last active measurement mode.

d

Store

Auto

Ident

S

1

LoBat

Tref25

nLF

/

K

%

Sal

°

AR

TP

C

M

m

%

mg/l

cm

S/

cm

cm

1/

°

F

ARng

Cal

Lin

Tref20

Sal

TDS

mbar

O

UpH

mV/pH

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Year

R

C A L

S E

RCL

CALIBRATION PROTOCOL
02.03.01 14:19
Device No.: 12345678

Calibration pH
Cal time: 01.10.01 / 15:20
Cal interval: 7d
AutoCal TEC Tauto
Buffer 1 2.00
Buffer 2 4.01
Buffer 3 7.00 *
Buffer 4 10.00
C1 174.1 mV 25.0°C
C2 -133.3 mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++

CALIBRATION 02
Cal time: 02.03.01 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++

CALIBRATION ISE
Cal time: 01.10.01 / 15:30
Std 1 10,00 mg/l
Std 2 20,00 mg/l
Std 3 50,00 mg/l