Note
When the AutoStore function is active at the same time, the data trans-
mission is performed according to the setting of the save interval (Int 1).
Set the save interval (Int 1) to OFF to activate the Data transmission in-
terval
(Int 2).

4.7.2

PC/external printer (RS232 interface)

Via the RS 232 interface, you can transmit the data to a PC or an ex-
ternal printer. Use the AK340/B (PC) or AK325/S (ext. printer) cable to
connect the interface to the instruments. The data output automatically
switches to the RS232 interface.

Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.

Set up the following transmission data on the PC/printer:

Socket assignment

4

Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur-
ing mode.

Baud rate

selectable between: 1200, 2400, 4800, 9600

Handshake

RTS/CTS + Xon/Xoff

Parity

none

Data bits

8

Stop bits

2

Cable length

Max. 15 m

1 CTS
2 RxD
3 Ground
4 TxD

1

2

3

4

RS 232