4 autoread ar (drift control), Autoread ar (drift control) – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 36

Advertising
background image

Operation

pH/ 3400i

36

4.4.4

AutoRead AR (Drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.

Criteria

With identical measurement conditions, the following applies:

1

Call up the measuring mode with <M> and/or <

> <> .

2

Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This mea-
sured value is transmitted to the interface.

4

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.

5

To terminate the AutoRead function: Press the <AR> key.

RCL

cm

cm

M

m

S/

%

/pH

mV

pH

U

TDS

Tref20

Cal

8

S

1

Sal

1/cmSal

AutoCal DIN

Lin

Tref25

nLF

AutoCal TEC

Auto

Store

LoBat

AR

Time

Baud

Day.Month No.

Ident

Year

mbar

°

F

/

K

%

6

7

1

9 2

°

C

TP

7

O

mg/l

ARng

Measured
parameter

Reproducibility

Response time

D. O. concentra-
tion

better than 0.05 mg/l

> 10 seconds

D. O. saturation
index

better than 0.6 %

> 10 seconds