3 general errors, General errors – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 66

Advertising
background image

What to do if...

pH/ 3400i

66

Measured value too low

Measured value too

high

6.3

General errors

Display LoBat

Instrument does not

react

to keystroke

Display

to

to

to

to

Probe symbol flashes

Message

StoFull

StoFull

StoFull

StoFull

Cause

Remedy

– Insufficient flow

– Provide flow to the probe

Cause

Remedy

– High amount of dissolved

substances

– Correct solubility function

using the salinity equivalent

– Air bubbles bump on the

membrane with high velocity

– Avoid direct flow to the

membrane

– The carbon dioxide pressure is

too high (> 1 bar)

– Measuring not possible

Cause

Remedy

– Batteries almost empty

– Exchange the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Switch the instrument on
while pressing the <CAL> key

Cause

Remedy

– Time-out of the interface

– Check the instrument that is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Output data storage and clear

data storage