2 oxi system messages, Oxi system messages – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 65

Advertising
background image

pH/ 3400i

What to do if...

65

6.2

Oxi system messages

Error message

0FL

0FL

0FL

0FL

Error message

E3

E3

E3

E3

Error message

E7

E7

E7

E7

AR flashes

continuously

Cause

Remedy

Display range exceeded

D. O. probe:

– Not connected

– Connect the probe

– Cable broken

– Replace probe

– Depleted

– Replace probe

– Short-circuit between gold and

lead electrode

– Clean probe and replace it if

necessary

Cause

Remedy

Invalid calibration

D. O. probe:

– Electrolyte solution depleted

– Regenerate probe

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate probe

– Worn out

– Replace probe

– Broken

– Replace probe

Cause

Remedy

Membrane damaged

– Membrane damaged

– Regenerate probe

– Membrane head not screwed

on tight enough

– Screw membrane head tight

Cause

Remedy

No stable measured value

– Membrane contaminated

– Clean membrane