4.4.6

Entering the salt content (salinity)

A salt content correction is required in the oxygen concentration mea-
surement of test samples with a salt content of more than 1 g/l. To do
this, you have to enter the salinity equivalent of the test sample (range
0.0 - 70.0) and to switch on the salinity correction.

Entering the salinity

Note
How to switch on the salt content correction is described on page 34.

1

Determine the salinity of the test sample.

2

Press the <CAL> key repeatedly until Sal appears on the dis-
play.

3

Enter the salt content with <

> <>.

4

Switch to the measuring mode with <M>.

°