11

Manual

US 3.10 713 Flow Converter 1210 - Sw 833062

MJK North America, Inc.
37 Sherwood Terrace, #126
Lake Bluff, Il 60044
Tel.: 847-482-8655
Fax: 847-482-8654
[email protected]
www.mjk.com

Measurement through a concrete deck with

45 - 60 ° edge cutoff.

Min. diameter (D) =

sensor spread

+ 50 %

Measurement through a pipe for foam protection.
when measuring on liquids prone to build up foam on the
surface, it can become necessary to measure through a
pipe, since the build up of foam rarely will occur inside the
pipe.

Measurement through a pipe for foam protection.

Measuring through a concrete deck
when the ultrasonic sensor is measuring through a con-
crete deck, the dimension of the opening should be made
as shown below: (See table 1 for the sensor spread.)

Min. diameter (D)

= sensor spread +

100 %

Measurement through a concrete deck with

sharp edge.

Min. diameter (D)

= sensor spread +

100 %