Actual level (0.5 ft)

Bottom of sensor

e.g. 4 ft.

Appendix E
System Package 2 - Shuttle

®

used as

signal amplifier
to ensure a correct flow measurement when using
a Shuttle

®

as signal amplifier, the mA signal from the

Shuttle

®

must correspond to the selected level range on

the MJK 713. Four parameters must therefore be set on
the Shuttle

®

: the measuring unit, the zero point, the level

range and the mA output zero/span.

the keyboard is used for the initial programming of the
Shuttle

®

and is therefore hidden behind the front lid. the

keys are marked with symbols indicating their function.

Display

LCD-display with

symbols for indication

during programming

and normal service.

Keyboard

Keyboard with a

brief instruction for

programming the

SONOLEV™.

The lower part can be

tipped up to show the

terminal designations.

when Shuttle

®

is connected to the power supply for the

first time, the following display will appear:

Press

(enter) once.

At the same moment Shuttle

®

registers an echo, the

zero point is automatically adjusted to the level that is
present in the flume or weir.
Furthermore, the mA output is set to 4 mA at the current
zero point and 20 mA at a level corresponding to a dis-
tance of 14 in from the ultrasonic sensor.


Setting of measuring unit

the measuring unit should be set to the same unit as
used on the MJK 713.

Press

(UnIt) once.

Select the measuring unit with the UP/Down keys and
confirm with enter.

Setting of level readout
the level readout should be set to the actual level in the
flume / weir.

Press

again.

Use the UP/Down keys and confirm with enter.