2b: We deliver two types of sensor brackets that can be used

in almost any installation. The bracket shown is a flume
mounting bracket (order no. 200230 / 200235).

3: the ultrasonic sensor should be installed so that

the bottom of the sensor is as close as possible to
the max. height of the head but not closer than the
blocking distance.

Measuring range:

0 - 1’

0 - 3’

Blocking distance: )*

16”

16”

Min. measuring range:

4”

12”

Max. measuring range:

12”

36”

Max. sensor height

:

30"

54"

)* 14” for System Package 2.

(w. Shuttle

®

Level Transmitter as signal amplifier.)

Maximum sensor height

= Blocking distance

+ measuring range

Blocking distance

Sensor height

Actual measurement

Measuring range