Manual – Xylem MJK 713 Open Channel Flowmeter User Manual
Page 8

8
Manual
US 3.10 713 Flow Converter 1210 - Sw 833062
MJK North America, Inc.
37 Sherwood Terrace, #126
Lake Bluff, Il 60044
Tel.: 847-482-8655
Fax: 847-482-8654
[email protected]
www.mjk.com
2b: We deliver two types of sensor brackets that can be used
in almost any installation. The bracket shown is a flume
mounting bracket (order no. 200230 / 200235).
3: the ultrasonic sensor should be installed so that
the bottom of the sensor is as close as possible to
the max. height of the head but not closer than the
blocking distance.
Measuring range:
0 - 1’
0 - 3’
Blocking distance: )*
16”
16”
Min. measuring range:
4”
12”
Max. measuring range:
12”
36”
Max. sensor height
➁
:
30"
54"
)* 14” for System Package 2.
(w. Shuttle
®
Level Transmitter as signal amplifier.)
Maximum sensor height
➁
= Blocking distance
➀
+ measuring range
➃
➀
Blocking distance
➁
Sensor height
➂
Actual measurement
➃
Measuring range
➀
➁
➂
➃