Setting of zero and span for the mA output

the mA output must be set to 4 mA at zero level (0.00 ft)
and 20 mA at max. level (= the max.
measuring range for MJK 713 - either 1 or 3 ft).

Press

once.

Use the UP/Down keys to set the zero point and confirm
with enter.

the mA output must be set to 20 mA at max. level
(1 or 3 ft).

Press

again.

Use the UP/Down keys to set the span and confirm with
enter.

Setting of relay output #1

If the mA output is looped through relay 1 as shown on
the connection diagram on page 14 (System Package 2),
MJK 713 will indicate a system error in case of a power
failure or a a system error on the Shuttle

®

:

Press

once.

Use the UP/Down keys to select SYSteM error func-
tion and confirm with enter:

Set the desired time delay for the relay:

Use the arrow keys and comfirm with enter.

Set the relay mode to ‘n.c’ - norMAllY CloSeD (‘n.c =
normally closed):

Use the arrow keys and comfirm with enter.

Shuttle

®

will now revert to normal read-out:

4 mA out

20 mA out

Zero flow level

Max. flow level