LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

Using the Defibrillator

TROUBLESHOOTING

This section explains problems you may encounter while using the LIFEPAK CR Plus or
LIFEPAK EXPRESS defibrillator. For information about keeping your defibrillator in a state of readiness,
refer to Section 5.

Table 3-1

Troubleshooting During Victim Use

Problem

Possible Cause

What To Do

CHECK PADS FOR GOOD
CONTACT

or

CHECK

CONNECTOR

voice

instructions are heard

Inadequate connection to the defibrillator •

Be sure that the connector is
completely inserted.

Electrode pads are not properly adhered
to the victim

Firmly press the pads on the
victim’s skin.

Clean, shave, and dry the
victim’s skin before placing
pads on skin.

Electrode pads are dry, damaged, or
have passed the expiration date

Replace the pads.

Electrode pads are not removed from the
blue plastic

Remove the pads from the
blue plastic and apply them
to the victim’s chest.

Defibrillator cannot
deliver the required
shock

Defibrillator internal battery power is low

Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.

Voice instructions sound
faint or distorted

Defibrillator internal battery power is low

Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.

MOTION DETECTED

and

STOP MOTION

voice

instructions are heard

Victim movement because of location

Move the victim to a stable
location, if possible.

Victim movement because of breathing

Stop CPR during analysis.

Check victim for normal
breathing.

Vehicle motion

Stop vehicle during analysis,
if possible.

Electrical/radio frequency interference

Move communication or
other suspected devices
away from the defibrillator
when possible.