Physio-Control LIFEPAK CR Plus User Manual

Page 41

Advertising
background image

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

5-3

©2009-2010 Physio-Control, Inc.

5

C

ari
ng f

o

r

Y

o

u

r Defi

bril

lator

Caring for Your Defibrillator

The replacement kit includes a CHARGE-PAK battery charger, one or two QUIK-PAK electrode packets,
and a CHARGE-PAK battery discharger. The discharger depletes a used battery charger so that it is
ready for recycling or disposal.

Note:

Always keep the CHARGE-PAK battery charger with its original defibrillator. When a

CHARGE-PAK battery charger is inserted into a defibrillator, it becomes internally linked to that
device. If the CHARGE-PAK battery charger is inserted into a second device, the device will not
function correctly. If this occurs, obtain a new replacement kit and replace both the CHARGE-PAK
battery charger and the QUIK-PAK electrodes at the same time.

Follow instructions provided in the replacement kit for battery charger and electrode packet
recycling/disposal.

WARNING!

Possible explosion or fire.

The CHARGE-PAK battery charger is not rechargeable. Do not attempt to recharge, open, crush, or
burn the battery, or it may explode or catch fire.

Advertising
This manual is related to the following products: