A-4

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

Specifications

User Interface

Environmental

Note:

All performance specifications defined assume that the unit has been stored (two hours

minimum) at the operating temperature prior to operation.

User Interface:

The user interface includes voice instructions, audible tones, and graphical
prompts.

Readiness Display

The readiness display shows the device status.

OK Indicator:

Indicates OK when the last self test was completed successfully. When the
OK indicator is visible, all other indicators are not visible.

The OK indicator is not displayed during device operation.

CHARGE-PAK
Indicator:

When displayed, replace the CHARGE-PAK battery charger.

Attention Indicator:

When first displayed, at least 6 discharges or 42 minutes of operating time
will remain.

Service Indicator:

Service required when displayed.

Operating Temperature: 0° to 50°C (32° to 122°F)

Storage Temperature:

-40° to 70°C (-40° to 158°F) with CHARGE-PAK battery charger and
electrodes, maximum exposure time limited to one week.

Atmospheric Pressure:

760 mmHg to 429 mmHg, 0 to 15,000 feet above sea level.

Relative Humidity:

5 to 95% (non-condensing)

Water Resistance:

IEC60529/EN60529 IPX4 “Splash proof” with electrodes connected and
CHARGE-PAK installed.

Shock:

MIL-STD-810E, Method 516.4, Procedure 1, (40g, 6 to 9 msec pulse, ½ sine
each axis).

Vibration:

MIL-STD-810E, Method 514.4, Helicopter – category 6 (3.75 g rms) and
Ground Mobile – category 8 (2.85 g rms).

EMC:

Refer to Appendix D for EMC information as defined in IEC 60601-1-2.