LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

Shock Advisory System

The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators were also tested using paced rhythms
recorded at high-fidelity from victims with implanted pacemakers. The high-fidelity pacemaker spikes
were also added to samples of ventricular fibrillation to test the defibrillator’s ability to reach a shock
decision in the case of ventricular fibrillation with an implanted, active pacemaker. The results are
summarized in the following table.

Table B-2

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance with Active Pacemakers

The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators were also tested using ECGs acquired
from hospitalized pediatric victims ranging in age from < 1 day old to 17 years old. The results are
summarized in the following table.

Nonshockable:
all other rhythms

531

>95% specificity

95.9% [94.5%]

Intermediate:
fine VF

29

Report only

96.6% [87.2%] sensitivity

1

From Physio-Control ECG database. Each sample is run 10 times asynchronously.

2

Association for the Advancement of Medical Instrumentation. DF39-1993 Standard for Automatic External Defibrillators

and Remote-Control Defibrillators. Arlington, VA: AAMI;1993.

3

Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting

Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety. AHA Task Force on
Automatic External Defibrillation, Subcommittee on AED Safety and Efficacy. Circulation, 1997, Vol. 95, 1677-1682.

4

LCL = 90% exact one-sided lower confidence limit

VF = ventricular fibrillation

VT= ventricular tachycardia

NSR = normal sinus rhythm

Rhythm Class

ECG Test
Sample Size

Performance Goal

Observed Performance

Shockable:
Coarse VF

35

>90% sensitivity

91.4% [81.9%]

Nonshockable:
Paced rhythms

35

>95% specificity

100.0% [93.6%]

Table B-3

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance Table for Pediatric ECGs

Rhythm Class

ECG Test

1

Sample Size

Performance Goal

2

Observed Performance

Sensitivity or Specificity [LCL]

3

Shockable:
coarse VF

90

>90% sensitivity

100.0% [97.5%]

Shockable:
shockable VT

11

>75% sensitivity

54.5% [31.8%]

Nonshockable:
NSR

424

>99% specificity

100.0% [99.5%]

Nonshockable:
asystole

95

>95% specificity

100.0% [97.6%]

Table B-1

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance Table for Adult ECGs (Continued)

Rhythm Class

ECG Test

1

Sample Size

Performance Goal

2,

3

Observed Performance
Sensitivity or Specificity [LCL]

4