Physio-Control LIFEPAK CR Plus User Manual

Page 53

Advertising
background image

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

A-1

©2009-2010 Physio-Control, Inc.

Ap
pe
nd
ixe
s

Specifications

SPECIFICATIONS

All specifications are at 20°C (68°F) unless otherwise stated.

Defibrillator

Waveform:

Biphasic Truncated Exponential, with voltage and duration compensation
for patient impedance. See Figure A-2, "Biphasic Waveform".

Patient Impedance Range: 7–300 ohms. If the detected impedance is outside this range, defibrillation

is inhibited and the user is prompted to connect electrodes.
The following specifications apply from 25 to 175 ohms.

Output Energy Sequence:

Multiple energy levels from 150 to 360 joules. Refer to Section 6,
Defibrillator Operating Settings.

Output Energy Accuracy:

10% of the energy setting into 50 ohms.
15% of the rated energy output into 25–175 ohms.

Rated energy output is the nominal delivered energy based on the energy
setting and patient impedance, as defined in the following chart.

*Energy setting selected

Figure A-1

Rated Energy Output

Rated Energy Output

En

er

g

y (

J)

Patient Impedance ()

Rated Energy Output

En

er

g

y (

J)

Patient Impedance ()