5-4

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

Caring for Your Defibrillator

Replacing the CHARGE-PAK Battery Charger

To replace the CHARGE-PAK Battery Charger:

1 Press

the

release latch (in the direction of the arrow) to

remove the used battery charger. The battery charger
springs outward from the defibrillator.

2

Insert the new battery charger into your defibrillator and
push until you hear it click into position.

3

Confirm that the

symbol disappears and that the OK

symbol appears in the readiness display.

Note:

If the

symbol appears after you replace the

battery charger, the internal battery is very low and needs
time to charge. It may take up to three days if you had the
defibrillator on for a long time or if you delivered many
shocks. The OK symbol appears when the internal battery
is charged.

Remember: If the defibrillator is needed for an emergency,
attempt to use it even if the

symbol is visible.

CAUTION!

Keep the defibrillator at temperatures between 0° – 50°C (32° – 122°F) while the new battery charger
charges the internal battery. The internal battery may not charge effectively at lower temperatures.
Temperatures exceeding 50°C (122°F) for longer than seven days can permanently damage the
internal battery.

Advertising
This manual is related to the following products: