LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

3-7

©2009-2010 Physio-Control, Inc.

3

U

sing the

Defibrilla

tor

Using the Defibrillator

Defibrillator does not
deliver voice
instructions or beeping
tones after you open it
(turn it on)

Depleted internal battery

•

Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.

•

Replace the CHARGE-PAK
battery charger as soon as
possible. After the

OK

symbol

appears on the readiness
display, return the defibrillator
to service.

•

Contact authorized service
personnel.

Speaker system failure

•

Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.

•

Contact authorized service
personnel.

The readiness display
is blank

The defibrillator has been turned on

•

Normal condition when the
defibrillator is in use.

Operating temperature is too low or too
high

•

Operate the defibrillator
within 0° to 50°C (32° to
122°F).

LCD not operating properly

•

Contact authorized service
personnel.

Table 3-1

Troubleshooting During Victim Use (Continued)

Problem

Possible Cause

What To Do