Electromagnetic Compatibility Guidance

Essential Performance

The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators maintain safe and effective performance of the
defibrillation therapy and patient monitoring functions when operated in the electromagnetic environment
specified in Tables 2 through 4.

Limitations Affecting Immunity to Electromagnetic Disturbances

The level of protection from electromagnetic disturbances is limited by several factors, including requirements
for protection from third-party defibrillators, patient safety isolation, and maintenance of adequate
signal-to-noise ratios for processing patient signals.

Table D-1

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use in the electromagnetic
environment specified below. The customer or the user of the defibrillator should ensure that the defibrillator is
used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions

CISPR 11

Group 1

The defibrillator uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.

RF emissions

CISPR 11

Class B

The defibrillator is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Not Applicable

Voltage fluctuations/
flicker emissions

IEC 61000-3-3

Not Applicable