Physical characteristics accessories – Physio-Control LIFEPAK CR Plus User Manual

Page 57

Advertising
background image

LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions

A-5

©2009-2010 Physio-Control, Inc.

Ap
pe
nd
ixe
s

Specifications

Physical Characteristics

Accessories

CHARGE-PAK Battery Charger

QUIK-PAK Electrode Pads

Height:

10.7 cm (4.2 in)

Width:

20.3 cm (8.0 in)

Depth:

24.1 cm (9.5 in), excluding handle

Weight:

2.0 kg (4.5 lb) with CHARGE-PAK and electrodes

Type:

Li/SO

2

Cl

2

Lithium Sulfuryl Chloride, 11.7V, 1.4 amp-hours.

Replacement:

Replace after each use or when CHARGE-PAK indicator is visible (typically
after 2 years)

Weight:

80.5 grams (0.18 lb)

Pads:

Pacing/defibrillation/ECG electrodes.

Pads Packaging:

User-intuitive, rapid-release QUIK-PAK electrodes allow the electrode pads
to be pre-connected to the device and protected under a top cover.

Pads Shelf Life:

Two years typical.

Electrode Shape:

Oval-rectangular.

Electrode Size:

11.2 cm (4.4 in) × 18.5 cm (7.3 in)

Lead Wire:

1.067 meters (3.5 feet)

Conductive Adhesive
Gel Contact Area:

82 cm

2

(12.8 in

2

)

Maximum Adhesion
Time:

24 hours

Maximum ECG
Monitoring Time:

24 hours

Maximum Number of
Defibrillation Pulses:

50 at 360 joules

Maximum Pacing
Duration:

Up to 12 hours

ECG is received from disposable defibrillation electrodes, standard placement (anterior-lateral), or
anterior/posterior placement.

Advertising